the ampacity of supply branch-circuit conductors and the current rating of overcurrent protective devices for diagnostic x-ray equipment shall not be less than percent of the momentary rating or percent of the long-time rating, whichever is greater.

Respuesta :

The ampacity of supply feeders and the current rating of overcurrent protective devices supplying two or more branch circuits supplying X-ray units shall not bet less than 50 percent of the momentary demand rating of the largest unit plus 25 percent of the momentary demand rating of the next largest unit plus 10 percent of the momentary demand rating of each additional unit.

  • The ampacity of supply branch-circuit conductors and the current rating of overcurrent protective devices shall not be less than 50 percent of the momentary-rating or 100 percent of the long-time rating, whichever is greater.
  • Where simultaneous biplane examinations are undertaken with the X-ray units, the supply conductors and overcurrent protective devices shall be 100 percent of the momentary demand rating of each X-ray unit.
  • The ampacity of conductors and rating of overcurrent protective devices shall not be less than 100 percent of the current rating of medical X-ray therapy equipment.